What Parafilm M is for, and what it will not do

Almost everyone who has worked in a laboratory has used it, and almost nobody has read its limits. They are short, and they decide whether it is the right closure for what you are about to do.

What it is

A blend of polyolefins and paraffin waxes, with no plasticisers in it. Odourless, colourless, semi-transparent and chemically inert. It carries no adhesive and needs no heat: pulled to over twice its length and pressed down, it grips itself as it recovers. The seal is the stretch.

That is also why it seals to itself rather than to glass. Wrap it past the rim and down the neck, so that the film has its own surface to close against.

The five limits

Heat accounts for three of them. The other two spoil an experiment rather than a seal.

Working range
−45 °C to +50 °C
Cold is not the problem. Warmth is.
Melting point
60 °C
A warm room is fine; a drying oven is not.
Autoclave
No
A cycle runs at 121 °C, twice the melting point. It will not survive it.
Gases
Permeable
Oxygen, carbon dioxide and water vapour pass through it.
Polar solvents
About 48 hours
It resists them for a working day or two, not indefinitely.

The one that catches people out

Moisture proof and airtight are not the same claim, and the film is only the first. It holds a liquid in and slows evaporation to almost nothing, which is what it is for. It does not hold an atmosphere: oxygen, carbon dioxide and water vapour cross it. Anything that depends on the headspace — anaerobic culture, a controlled gas, a long incubation where slow exchange matters — wants a stopper, not a film.

Which width

Width is the only real decision. Four inches closes a flask neck or a beaker in a single wrap and is what most benches keep. Two inches is for tubes, plates and small joints, and a roll of it goes about twice as far because you are not throwing half of each piece away.

Length follows from that rather than the other way round. See what we stock →

Questions we are asked

Can you autoclave Parafilm M?
No. Parafilm M melts at 60 °C and an autoclave cycle runs at 121 °C. Seal vessels after sterilisation, not before.
What temperature can Parafilm M take?
The manufacturer's stated working range is −45 °C to +50 °C. It stays usable in a freezer; it softens and then melts with heat, at 60 °C.
Is Parafilm M airtight?
No. It is a barrier against evaporation and volume loss, but it is permeable to oxygen, carbon dioxide and water vapour. For a controlled headspace or anaerobic work it is the wrong closure.
How far does Parafilm M stretch?
Over 200% of its original length. The stretch is what makes the seal: it grips itself as it recovers, with no adhesive and no heat.

The figures here are the manufacturer's and its distributors', restated rather than improved — Amcor, who make it. We are an authorized reseller: what we can tell you about the film is what they publish about it, and we would rather say so than round it off.